Vedere la pagina 98
Programming Manual Lx \ Ls Series II
95
VOLTage:REFerance
This command select the referance Voltage. The following voltage refernace can be selected:
INTernal This is the internal voltage referance for programmable voltage
EXTernal This is the external DC voltage referance input to program the output
voltage. 0 to +10 voltage represents 0 to full scale output voltage. This is
the RPV option.
Command Syntax [SOURce:]VOLTage:Refernace <source>
Parameters INTernal | EXTernal
*RST Value INTernal
Examples VOLT:REF EXT
Query Syntax [SOURce:]VOLTage:REFerance?
Returned Parameters <CRD>
Related Commands VOLT:ALC:SOUR OFF
Note: The VOLT:ALC:SOUR command must be set to off prior to selection of the referance to
external.