Programming Manual Lx \ Ls Series II
108
Command Syntax SYSTem:COMMunicate:SERial:BAUD <NRF>
Parameters 9600 | 19200 | 38400 | 57600 | 115200 | 230400 | 460800
Example SYST:COMM:SER:BAUD 38400
Returned Parameters <NR1>
SYSTem:COMMunicate:SERial:PARity
This command will set the SERial communication parity.
Command Syntax SYSTem:COMMunicate:SERial:PARity <parity>
Parameters NONE|EVEN|ODD
Example SYST:COMM:SER:PAR NONE
Returned Parameters <CRD>
SYSTem:COMMunicate:SERial:BITS
This command will set the SERial communication number of data bits.
Command Syntax SYSTem:COMMunicate:SERial:BITS <NRF>
Parameters 7|8
Example SYST:COMM:SER:BITS 8
Returned Parameters <NR1>
SYSTem:COMMunicate:SERial:SBITs
This command will set the SERial communication number of stop bits.
Command Syntax SYSTem:COMMunicate:SERial:SBITs <NRF>
Parameters 1|2
Example SYST:COMM:SER:SBITs 1
Returned Parameters <NR1>
SYSTem:CONFiguration
This query format of this command returns the installed options and configuration settings of the
power source. It is not possible to change the configuration as only the query form is supported.
Command Syntax SYSTem:CONFigure?
Query Syntax SYSTem:CONFigure?
Returned Parameters <AARD>
Related Commands *OPT?
SYSTem:CONFiguration:NOUTputs
This command selects the number of outputs phases for the AC source. This requires that the AC
source is capable of switching between single and three-phase mode.
Note: Execution of this command disables all outputs, clears lists and *RCL states to the
initialization default values, reconfigures current readback and programming calibration
constants.
Once configured, the instrument behaves transparently as either a single-phase source or as a
three phase source depending on the selected configuration.