Programming Manual Lx \ Ls Series II
117
Command Syntax TRIGger:SEQuence3[:IMMediate]
TRIGger:ACQuire[:IMMediate]
Parameters None
Examples TRIG:ACQ TRIG:SEQ3:IMM
Related Commands ABOR TRIG:SOUR TRIG:DEL TRIG:SYNC
TRIG:SYNC:PHAS INIT INIT:CONT *TRG *WAI
TRIGger:SEQuence3:SOURce
TRIGger:ACQuire:SOURce
These commands select the trigger source for a triggered measurement sequence as follows:
BUS IEEE-488 device, *TRG, or <GET> (Group Execute Trigger)
EXTernal backpanel Trigger In1 SMA connector
TTLTrg the signal driving the Trigger Out1 SMA
Command Syntax TRIGger:SEQuence3:SOURce<CRD>
Command Syntax TRIGger:ACQuire:SOURce<CRD>
Parameters BUS | EXT | TTLT
*RST Value BUS
Examples TRIG:ACQ:SOUR BUS
TRIG:SEQ3:SOUR EXT
Query Syntax TRIGger:SEQuence3:SOURce?
TRIGger:ACQuire:SOURce?
Returned Parameters <CRD>
Related Commands ABOR TRIG:SOUR TRIG:DEL TRIG:SYNC
TRIG:SYNC:PHAS INIT INIT:CONT *TRG *WAI
TRIGger:SEQuence1:DEFine
TRIGger:SEQuence2:DEFine
TRIGger:SEQuence3:DEFine
These commands define the names that are aliased to trigger sequences 1, 2 and 3. The
command accepts only TRANsient for sequence 1, SYNChronous for sequence 2, and ACQuire
for sequence 3 as predefined names. The query allows the user to query the instrument names
aliased to sequences 1, 2, and 3.
Command Syntax TRIGger:SEQuence1:DEFine TRANsient
TRIGger:SEQuence2:DEFine SYNChronous
TRIGger:SEQuence3:DEFine ACQuire
Parameters TRANsient, SYNChronous, ACQuire
Examples SEQ1:DEF TRAN SEQ3:DEF ACQ
Query Syntax TRIGger:SEQuence1:DEFine?
TRIGger:SEQuence2:DEFine?
TRIGger:SEQuence3:DEFine?
Returned Parameters <CRD>