Programming Manual Lx \ Ls Series II
119
5.1 *CLS
This command clears the following registers (see chapter 7 under Programming the Status and
Event Registers for descriptions of all registers):
Standard Event Status
Operation Status Event
Questionable Status Event
Status Byte
Error Queue
Command Syntax *CLS
Parameters None
*ESE
This command programs the Standard Event Status Enable register bits. The programming
determines which events of the Standard Event Status Event register (see *ESR?) are allowed to
set the ESB (Event Summary Bit) of the Status Byte register. A "1" in the bit position enables the
corresponding event. All of the enabled events of the Standard Event Status Event Register are
logically ORed to cause the Event Summary Bit (ESB) of the Status Byte Register to be set. See
section 7.5 for descriptions of the Standard Event Status registers.
The query reads the Standard Event Status Enable register.
Bit Position
7
6
5
4
3
2
1
0
Bit Name
PON
not used
CME
EXE
DDE
QYE
not used
OPC
Bit Weight
128
32
16
8
4
1
Table 5-1: Bit Configuration of Standard Event Status Enable Register
CME
Command error
DDE
Device-dependent error
EXE
Execution error
OPC
Operation complete
PON
Power-on
QYE
Query error
Command Syntax *ESE <NRf>
Parameters 0 - 255
Power-On Value (See *PSC)
Example *ESE 129
Query Syntax *ESE?
Returned Parameters <NR1>(Register value)
Related Commands *ESR? *PSC *STB?
5.2 *ESR?
This query reads the Standard Event Status Event register. Reading the register clears it. The bit
configuration of this register is the same as the Standard Event Status Enable register (see
*ESE). See section 7.5 for a detailed explanation of this register.
Vedere la pagina 122
1 2 ... 118 119 120 121 122 123 124 125 126 127 128 ... 241 242

Commenti su questo manuale

Nessun commento