Programming Manual Lx \ Ls Series II
96
VOLTage:SENSe:SOURce
These commands select the source from which the output voltage is sensed. The following
voltage sense sources can be selected:
INTernal This senses the voltage at the output of the power amplifier on the
inboard side of the output disconnect relay.
EXTernal This senses the output voltage at the user's sense terminals, which
allows remote voltage sensing at the load.
Command Syntax [SOURce:]VOLTage:SENSe:SOURce<source>
[SOURce:]VOLTage:ALC:SOURce<source>
Parameters INTernal | EXTernal
*RST Value INTernal
Examples VOLT:SENS:SOUR INT
Query Syntax [SOURce:]VOLTage:SENSe:SOURce?
Returned Parameters <CRD>
Related Commands VOLT:ALC:SOUR EXT
Note: The VOLT:ALC:SOUR command is an alias for the VOLT:SENS:SOUR command. Both
perform the same function. This is done for backward compatability with the Agilent 6834B.
VOLTage:SLEW
This command sets the slew rate for all programmed changes in the ac rms output voltage level
of the AC source. A parameter of MAXimum or INFinity will set the slew to its maximum possible
rate. The SCPI representation for INFinity is 9.9E37. This command does not affect the rate at
which programmed dc offset changes occur.
Command Syntax [SOURce:]VOLTage:SLEW[:IMMediate]<NRf+>|INFinity
Parameters 1E-3 to 9.9E37 | INFinity |MINimum | MAXimum
Unit V/S (volts per second)
*RST Value INFinity
Examples VOLT:SLEW 1 VOLT:SLEW MAX VOLT:SLEW INF
Query Syntax [SOURce:]VOLTage:SLEW[:IMMediate]?
Returned Parameters <NR3>
Related Commands VOLT:SLEW:MODE VOLT:SLEW:TRIG