Programming Manual Lx \ Ls Series II
64
OUTPut:PROTection:DELay
This command sets the delay time between the programming of an output change that produces
a CL or UNREG status condition and the recording of that condition by the Status Operation
Condition register. The delay prevents momentary changes in status that can occur during
programming from being registered as events by the status subsystem. In most cases these
temporary conditions are not considered an event, and to record them as such would be a
nuisance.
Command Syntax OUTPut:PROTection:DELay<NRf>
Parameters 0 to 32 | MINimum | MAXimum
Unit S (seconds)
*RST Value 100 milliseconds
Examples OUTP:PROT:DEL 75E-1
Query Syntax OUTPut:PROTection:DELay?
Returned Parameters <NR3>
Related Commands OUTP:PROT:CLE *RCL *SAV
OUTPut:RI[:LEVel]
This command sets the remote inhibit level mode. Factory default is LOW, which requires a
contact closure to open the output relay. The level can be reversed by setting it to HIGH. Once
set, the RI level setting is retained each time the power source is powered up. Note that this
command is only implement with firmware revision 1.69 or higher. Lower firmware revisions only
provide the default LOW setting.
Command Syntax OUTPut:RI:LEVel
Parameters LOW | HIGH
*RST Value LOW
Examples OUTP:RI:LEV HIGH
Query Syntax OUTP:RI:LEV?
Returned Parameters <CRD>
Related Commands OUTP OUTP:RI:MODE
OUTPut:RI:MODE
This command selects the mode of operation of the Remote Inhibit protection. The following
modes can be selected:
LATChing A TTL low at the RI input latches the output in the protection shutdown
state, which can only be cleared by OUTPut:PROTection:CLEar.
LIVE The output state follows the state of the RI input. A TTL low at the RI
input turns the output off; a TTL high turns the output on.
OFF The instrument ignores the RI input.
The RI output state is saved as part of an instrument setup using the *SAV command. It can be
made part of the power on setting if needed. The default state is LIVE.
Command Syntax OUTPut:RI:MODE <mode>
Parameters LATChing | LIVE | OFF
*RST Value OFF
Examples OUTP:RI:MODE LIVE
Query Syntax OUTPut:RI:MODE?
Returned Parameters <CRD>
Related Commands OUTP:PROT:CLE OUTP:RI:LEV