Programming Manual Lx \ Ls Series II
88
4.21 Source Subsystem - Pulse
This subsystem controls the generation of output pulses. The PULSe:DCYCle, PULSe:HOLD,
PULSe:PERiod, and PULSe:WIDTh commands are coupled, which means that the values
programmed by any one of these commands can be affected by the settings of the others. Refer
to the tables under PULSe:HOLD for an explanation of how these commands affect each other.
Subsystem Syntax
[SOURce:]
PULSe
:COUNt <n> | INFinity Selects transient pulse count
:DCYCle <n> Selects pulse duty cycle
:HOLD WIDTh |DCYCle Selects parameter that is held constant
:PERiod <n> Selects pulse period when the count is greater than 1
:WIDTh <n> Selects width of the pulses
PULSe:COUNt
This command sets the number of pulses that are output when a triggered output transient
occurs. The command accepts parameters in the range 1 through 2E8. If INFinity or MAXimum is
sent, the output pulse repeats indefinitely.
Command Syntax [SOURce:]PULSe:COUNt<NRf+> | INFinity
Parameters 1 to 2E8 | MINimum | MAXimum | INFinity
*RST Value 1
Examples PULS:COUN 3 PULS:COUN MIN PULS:COUN INF
Query Syntax [SOURce:]PULS:COUNt?
Returned Parameters <NR3>
Related Commands PULS:DCYC PULS:HOLD PULS:PER PULS:WIDT
PULSe:DCYCle
This command sets the duty cycle of the triggered output pulse. The duty cycle units are specified
in percent.
Command Syntax [SOURce:]PULSe:DCYCle<NRf+>
Parameters 0 to 100%|MINimum|MAXimum
*RST Value 50%
Examples PULS:DCYC 75 PULS:DCYC MAX
Query Syntax [SOURce:]PULSe:DCYCle?
Returned Parameters <NR3>
Related Commands PULS:COUN PULS:HOLD PULS:PER PULS:WIDT
PULSe:HOLD
This command specifies whether the pulse width or the duty cycle is to be held constant when the
pulse period changes. The following tables describe how the duty cycle, period, and width are
affected when one, two, or all three parameters are set in a single program message.
Command Syntax [SOURce:]PULSe:HOLD<parameter>
Parameters WIDTh|DCYCle
*RST Value WIDTh
Examples PULS:HOLD DCYC
Query Syntax [SOURce:]PULSe:HOLD?
Returned Parameters <CRD>
Related Commands PULS:COUN PULS:DCYC PULS:PER PULS:WIDT