Programming Manual Lx \ Ls Series II
182
8.2.16 End of String Delimiter
The End of String (EOS) delimiter recognized by the Ls AC Power source in APE mode is the ASCII
Line Feed (LF). Carriage Return (CR) followed by Line Feed may also be used for EOS. The End
or Identify (EOI) IEEE-488 message END will also be recognized. The END message is sent by
setting the IEEE-488 End or Identify line true with the last data byte.
8.2.17 Group Execute Trigger - GET
The trigger mode is enabled when the mnemonic TRG is added to a setup string. The trigger
command may be inserted anywhere in the string. When the mnemonic is detected, it will delay
execution of the new setup values until the GPIB Device Trigger is sent by the bus controller.
A GPIB Device Trigger will also terminate a programmed ramp or other program.
The following setup string will recall the values from register 9 and delay execution until the GET
message is received. (Note: GET is the abbreviation for the GPIB Group Execute Trigger message
and does not represent a series of ASCII characters.
REC9 TRG