Programming Manual Lx \ Ls Series II
38
Command Syntax CALibrate:MEASure:VOLTage <Nrf>
Parameters Reference instrument voltage reading
Query Syntax CALibrate:MEASure:VOLTage?
Returned Parameters <NR3>
Examples CAL:MEAS:VOLT
Related Commands CAL:SAVE CAL:MEAS:CURR
CALibrate:PASSword
This command can only be used to unlock the calibration mode. Once unlocked, non-query
calibration commands will be accepted. Query commands are always accepted.
Command Syntax CALibrate:PASSword<NRf>
Parameters <high voltage range> (default)
Examples CAL:PASS 300
Related Commands none
CALibrate:SAVE
This command can only be used in calibration mode. It saves any new calibration constants (after
a current or voltage calibration procedure has been completed) in nonvolatile memory.
Command Syntax: CALibrate:SAVE
Parameters None
Examples CAL:SAVE
Related Commands CAL:CURR CAL:VOLT
CALibrate[:SOURce]:PHASe
Phase Selectable
This command can be used to set the phase calibration offset coefficient. Use the INST:SEL or
INST:NSEL to select the desired phase. This allows the phase for voltage 2 and 3 (B and C) to
be adjusted with respect to phase A. The query format returns the actual calibration coefficient
Command Syntax CALibrate[:SOURce]:PHASe <Nrf>
Parameters <NRf+>
Query Syntax CALibrate[:SOURce]:PHASe?
Returned Parameters <NR3>
Examples CAL:PHAS 1.3
Related Commands none
CALibrate[:SOURce]:VOLTage:HRANge
Phase Selectable
This command can be used to query the output voltage calibration coefficient. Only the query
format is available. Use the INST:SEL or INST:NSEL to select the desired phase. The parameter
can be any integer value between 0 and 4000. Factory default is 450.
Command Syntax CALibrate[:SOURce]:VOLTage:HRANge <Nrf>
Parameters 0 to 4000
Query Syntax CALibrate[:SOURce]:VOLTage:HRANge?
Returned Parameters <NR3>
Examples CAL:VOLT:HRAN 450
Related Commands none