Programming Manual Lx \ Ls Series II
82
Command Syntax [SOURce:]LIST:MODE
Parameters BOT | EOT
Examples LIST:MODE EOT
Query Syntax [SOURce:]LIST:MODE?
Returned Parameters <CRD>
Related Commands LIST:COUN LIST:DWEL LIST:STEP
LIST:PHASe
Phase Selectable
This phase selectable command sets the sequence of phase list points. The phase points are
given in the command parameters, which are separated by commas. The order in which the
points are entered determines the sequence in which they are output when a list is triggered.
Changing list data while a subsystem is in list mode generates an implied ABORt.
Command Syntax [SOURce:]LIST:PHASe<NRf+>{,<NRf+>}
Parameters 360 through +360
Examples LIST:PHAS 90,120,135
Query Syntax [SOURce:]LIST:PHAS?
Returned Parameters <NR3>
Related Commands LIST:PHAS:POIN? LIST:COUN LIST:DWEL LIST:STEP
LIST:PHASe:POINts?
This query returns the number of points specified in LIST:PHASe. Note that it returns only the total
number of points, not the point values.
Query Syntax SOURce:]LIST:PHASe:POINts?
Returned Parameters NR3>
Example IST:PHAS:POIN?
Related Commands IST:FREQ LIST:DWEL
LIST:SHAPe
This command sets the sequence of the waveform shape entries. The order in which the shapes
are given determines the sequence in which the list of shape will be output when a list transient is
triggered. Changing list data while a subsystem is in list mode generates an implied ABORt. The
following shapes may be specified:
SINusoid A sinewave is output
SQUare A squarewave is output
CSINusoid The output is a clipped sine waveform. Both positive and negative peak
amplitudes are clipped at a value determined by the
SOURce:FUNCtion:SHAPe:CSINusoid setting.
<waveform_name> The output shape is described by one of the user-defined waveform
tables.
The maximum peak voltage that the AC source can output is 425 V peak. This includes any
combination of voltage and function shape values. Therefore, the maximum value that can be
programmed depends on the peak-to-rms ratio of the selected waveform. For a sinewave, the
maximum voltage that can be programmed is 300 V rms.
Note: You cannot program a voltage that produces a higher volt-second on the output than a
300V rms sinewave.