BPS / MX / RS Series SCPI Programming Manual AMETEK Programmable Power
Manual P/N 7003-961 Rev. AA 89
SENSe:SWEep:TINTerval
This command and its query format allow setting the time period between samples when
voltage and current digitization is controlled by the acquire trigger sequence. The
query response of the sample period query will be:
MX Series I: 25.6 to 256 μsec for single phase and 76.8 to 768 μsec for three phase AC/DC
source configurations respectively. The value of TINT must be set in increments of 25 or 75
μsec. The controller will round to the nearest available multiple of 25.6 or 76.8 μsec.
MX Series II and RS Series: 10.4 to 104 μsec for single phase and 31.2 to 312 μsec for
three phase AC/DC source configurations respectively. The value of TINT must be set in
increments of 10.4 or 31.2 μsec. The controller will round to the nearest available multiple of
10.4 or 31.2 μsec.
Command Syntax SENSe:SWEep:TINTerval <NRf+>
Parameters MX Series I: 25.6 to 256 for single phase configuration
76.8 to 768 for three phase configuration
MX/RS Series II:10.4 to 104 for single phase configuration
31.2 to 312 for three phase configuration
*RST Value MX Series I: 25.6 or 76.8
MX/RS Series II:10.4 or 31.2
Examples SENS:SWE:TINT-150
Query Syntax SENSe:SWEep:TINTerval?
Returned Parameters <NR2>
Related Commands SENS:SWE:OFFS MEAS:ARR