FREQuency:SLEW:TRIGgered
This command sets the rate at which frequency changes during a triggered output transient.
Instantaneous frequency changes can be obtained by sending MAXimum.
Command Syntax [SOURce:]FREQuency:SLEW:TRIGgered <NRf+>
Parameters 0.01 to 1E9⏐MAXimum
Unit Hz (Hertz per second)
*RST Value MAXimum
Examples FREQ:SLEW:TRIG 75 FREQ:SLEW:TRIG MAX
Query Syntax [SOURce:]FREQuency:SLEW:TRIG?
Returned Parameters <NRf>
Related Commands FREQ:SLEW:MODE FREQ
FREQuency:TRIGgered
This command programs the frequency that the output will be set to during a triggered step
or pulse transient.
Command Syntax [SOURce:]FREQuency:TRIGgered <NRf+>
Parameters 0 to maximum frequency range specified by the LIMit:FREQuency
command
Unit Hz (Hertz)
*RST Value 60 Hz
Example FREQ:TRIG 50
Query Syntax [SOURce:]FREQuency:TRIGgered?
Returned Parameters <NR2>
Related Commands FREQ FREQ:MODE