AMETEK Programmable Power BPS / MX / RS Series SCPI Programming Manual
102 Manual P/N 7003-961 Rev AA
PONSetup:OUTPut[:RELay]
This command sets the initial state of the output relay at the power on.
Command Syntax [SOURce:]PONSetup:OUTPut[:RELay <boolean>
Parameters 0|1|OFF|ON
Examples OUTP ON OUTP 0
Query Syntax PONS:OUTP?
Returned Parameters 0 | 1
PONSetup:PHASe[:ANGLe]
This command sets the initial phase angle of the power source relative to external reference.
Command Syntax [SOURce:]PONSetup:PHASe[:ANGLe <NRf+>
Parameters ± 360
Examples PONS:PHAS 0
Query Syntax PONS:PHAS?
Returned Parameters <NR2>
PONSetup:SENSe
This command determines the source of its voltage sense at the power on. if internal, the
source uses its connection to sense its output voltage. if external, the sense connection
must be connected at the load.
Command Syntax [SOURce:]PONSetup:SENSe <source>
Parameters INTernal|EXTernal
Examples PONSetup:SENSe INT
Query Syntax PONSetup:SENS?
Returned Parameters <CRD>
Related Commands VOLTage:SENSe
PONSetup:VOLTage[:LEVel]
This command sets the output voltage level at the power on. This voltage level can not
exceed the limit at the initial voltage range.
Command Syntax [SOURce:]PONSetup:VOLTage[:LEVel] <NRf+>
Parameters set by the LIMit:VOLTage and voltage mode AC or DC
Examples PONS:VOLT 0
Query Syntax PONS:VOLT?
Returned Parameters <NR2>