AMETEK Programmable Power BPS / MX / RS Series SCPI Programming Manual
174 Manual P/N 7003-961 Rev AA
IEC411:DIPS:RUN[:ALL]
This command will cause the automated test sequence by the standard to run. Refer to
Table 13 in the User Manual.
Command Syntax IEC411:DIPS:RUN[:ALL]
Parameters <none>
Examples IEC411:DIPS:RUN
Query Syntax none
Returned Parameters <none>
IEC411:DIPS:RUN:SINGle
This command will run a single test once. The test is defined by the dip parameters voltage,
cycle and angle.
Command Syntax IEC411:DIPS:RUN:SINGle
Parameters <none>
Examples IEC411:DIPS:RUN:SING
Query Syntax none
Returned Parameters <none>
IEC411:DIPS:VOLTage[:PERCent]
This command defines the dip voltage as a percentage of the nominal voltage when the run
single command is used.
Command Syntax IEC411:DIPS:VOLTage[:PERCent]
Parameters <numeric>
Examples IEC411:DIPS:VOLT 20
Query Syntax IEC411:DIPS:VOLT?
Returned Parameters <NR2>
IEC411:DIPS:CYCLes
This command defines the duration of the dip voltage expressed in the number of cycles.
Command Syntax IEC411:DIPS:CYCLes
Parameters <numeric>
Examples IEC411:DIPS:VOLT 20
Query Syntax IEC411:DIPS:VOLT?
Returned Parameters <NR2>