Ametek BPS Series Programming Manual Manuale Utente Pagina 105

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 226
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 104
BPS / MX / RS Series SCPI Programming Manual AMETEK Programmable Power
Manual P/N 7003-961 Rev. AA 105
Parameter Set Action
DCYCle PERiod WIDTh
Sets WIDTh. If WIDTh < PERiod, recalculates DCYCle;
otherwise, recalculates the PERiod and DCYCle.
Sets PERiod. If WIDTh < PERiod, recalculates DCYCle;
otherwise, recalculates the PERiod and DCYCle.
Sets DCYCle and recalculates PERiod
Table 4-1 : PULSe:HOLD = WIDTh parameters
Parameter Set Action
DCYCle PERiod WIDTh
Sets WIDTh and recalculates the PERiod
Sets PERiod and recalculates the WIDTh
Sets DCYCle and recalculates PERiod
Table 4-2 : PULSe:HOLD = DCYCle parameters
PULSe:DCYCle
This command sets the duty cycle of the triggered output pulse. The duty cycle units are
specified in percent.
Command Syntax [SOURce:]PULSe:DCYCle<NRf+>
Parameters 0 to 100%|MINimum|MAXimum
*RST Value 50%
Examples PULS:DCYC 75
PULS:DCYC MAX
Query Syntax [SOURce:]PULSe:DCYCle?
Returned Parameters <NR2>
Related Commands PULS:COUN PULS:PER PULS:WIDT
PULSe:PERiod
This command sets the period of a triggered output transient The command parameters are
model-dependent.
Command Syntax [SOURce:]PULSe:PERiod<NRf+>
Parameters 2 msecs to 90,000 secs|MINimum|MAXimum
Unit S (seconds)
*RST Value 1 sec
Examples PER 0.001PER MIN
Query Syntax [SOURce:]PERiod?
Returned Parameters <NR2>
Related Commands PULS:COUN PULS:DCYC PULS:WIDT