BPS / MX / RS Series SCPI Programming Manual AMETEK Programmable Power
Manual P/N 7003-961 Rev. AA 47
4.2.7 Output - AC Voltage
CALibrate[:SOURce]:VOLTage[:AC]:LRANge[:FSCale] <NRf+>
This command will result in the calibration coefficent being calculated for the AC full scale
output voltage at the low voltage range.
Command Syntax CALibrate:VOLTage:LRANge <NRf+>
Parameters <NRf> (programmed output voltage)
Examples CAL:VOLT:LRAN 120
Query Syntax CALibrate:VOLTage:LRANge?
Returned Parameters <NRf> (cal coefficient value range 0.9 to 1.1)
Related Commands CAL:VOLT:LRAN:ZERO CAL:VOLT:LRAN:HFR
CALibrate[:SOURce]:VOLTage[:AC]:LRANge:ZERO <NRf+>
This command will set the calibration coefficent for the output voltage offset at the low
voltage range.
Command Syntax CALibrate:VOLTage:LRANge:ZERO <NRf+>
Parameters <NRf> (a value between -127 and +128)
Examples CAL:VOLT:LRAN:ZERO +10
Query Syntax CALibrate:VOLTage:LRANge:ZERO?
Returned Parameters <NR1> (value range -127 to +128)
Related Commands CAL:VOLT:LRAN CAL:VOLT:LRAN:HFR
CALibrate[:SOURce]:VOLTage[:AC]:LRANge:HFRequency <NRf+>
This command will set the calibration coefficent for the AC full scale output voltage at the low
voltage range and at high output frequency
Command Syntax CALibrate:VOLTage:LRANge:HFRequency <NRf+>
Parameters <NRf> (a value between -127 and +128)
Examples CAL:VOLT:LRAN:HFRequency +10
Query Syntax CALibrate:VOLTage:LRANge:HFRequency?
Returned Parameters <NR1> (value range -127 to +128)
Related Commands CAL:VOLT:LRAN CAL:VOLT:LRAN:ZERO
CALibrate[:SOURce]:VOLTage[:AC]:HRANge[:FSCale] <NRf+>
This command will result in the calibration coefficent being calculated for the AC full scale
output voltage at the high voltage range.
Command Syntax CALibrate:VOLTage:HRANge <NRf+>
Parameters <NRf> (programmed output voltage)
Examples CAL:VOLT:HRAN -2
Query Syntax CALibrate:VOLTage:HRANge?
Returned Parameters <<NRf> (cal coefficient value range 0.9 to 1.1)
Related Commands CAL:VOLT:HRAN:ZERO CAL:VOLT:HRAN:HFR