DLM-E Series and DCS-E Series Isolated Analog Programmer Operation
2.3.3 Voltage Source Programming of OVP
SW1 Switch Settings
SW1-3 ON (1) = 0–5V
SW1-3 OFF(0) = 0–10V
2361213 1
25
1416
Set up for voltage source programming of OVP as follows:
1. Set SW1-3, OVP PROG, to ON (1) for 0-5VDC programming range.
2. Set SW1-3, OVP PROG, to OFF (0) for 0-10VDC programming range.
3. Connect the external programming voltage source to the ISOLATED ANALOG
PROGRAMMER connector, J4, with positive to Pin-5 and negative to Pin-12.
4. Connect Pin-1, ANALOG-CONTROL, of the ISOLATED ANALOG PROGRAMMER
connector, J4, to Pin-2 to enable remote control.
5. Connect Pin 16, OVP RESISTANCE PROGRAMMING OUTPUT, of the ISOLATED
ANALOG PROGRAMMER connector, J4, to Pin 6 to disable the default OVP
programming connection.
6. Program the other parameters to the desired limit values.
M51A Option 2-9